Differently zoomed images of same device requires individual calibration for accurate measurement of object in an images.how to manage a single calibration on such images ?
1 visualizzazione (ultimi 30 giorni)
Mostra commenti meno recenti
vaishali
il 18 Ott 2013
Commentato: Image Analyst
il 18 Ott 2013
ultrasound images of same device differently zoomed and saved for further processing
0 Commenti
Risposta accettata
Image Analyst
il 18 Ott 2013
You can't. Different zoom levels will have a different calibration scale factor and there is no way around that if you want accurate lengths and areas out.
2 Commenti
Image Analyst
il 18 Ott 2013
Modificato: Image Analyst
il 18 Ott 2013
Yes, of course. If you recalibrate after you change the field of view then the measurements will be accurate.